Fully Automatic Control Panel The Fully Automatic Control Panel is designed to supply medical gas at a constant pressure from either bank of cylinders in to the MGPs. The unit shall have separate pressure regulators for each bank of cylinders and control system that shall be designed for the case of maintenance. Cylinders can be changed or the pressure regulators removed for overhaul without loss of continuity of the gas supply. The system shall be provided with monitoring to detect and display.
- Running bank operating
- Running bank low
- Reserve Bank operating
- Reserve bank low
- Pipeline pressure fault
The normal operation of the changeover and control depends on an electrical supply. The design shall ensure that in the event of electrical supply failure there is no disruption to the flow of gas into the MGPS.
|